Tympanometer: Practical Guide

Tympanometer Guide: How to Use, Types & Safety | PediaDevices
Essential Device for Testing Middle Ear Function in Children and Adults

Introduction

A tympanometer is a medical device used to test the function of the middle ear and eardrum (tympanic membrane). It measures how well the eardrum moves in response to changes in air pressure. This test, called tympanometry, helps detect fluid in the middle ear, ear infections, eustachian tube problems, and other ear conditions. The test is quick, painless, and commonly used in children to diagnose ear problems.

Key Point: Tympanometry is not a hearing test. It checks middle ear health, while hearing tests measure how well a person can hear sounds.

Purpose and Uses

What Does a Tympanometer Measure?

  • Movement of the eardrum in response to air pressure changes
  • Pressure inside the middle ear space
  • Fluid presence behind the eardrum
  • Eustachian tube function
  • Eardrum perforation or scarring

Common Uses

  • Diagnosing middle ear infections (otitis media)
  • Detecting fluid in the middle ear (otitis media with effusion)
  • Checking for eardrum perforation
  • Evaluating eustachian tube dysfunction
  • Monitoring ear conditions before and after treatment
  • Screening children with speech or language delays
  • Pre-surgical assessment for ear tubes

Where Are Tympanometers Used?

  • Pediatric clinics and hospitals
  • Ear, Nose, and Throat (ENT) specialist offices
  • Audiology centers
  • Primary care clinics
  • School health programs
  • Community health centers

Types of Tympanometers

Type Features Best For
Screening Tympanometer Basic pass/fail results, portable, simple operation Quick screening in clinics, schools
Diagnostic Tympanometer Detailed graphs, multiple test frequencies, comprehensive data ENT specialists, audiologists, detailed diagnosis
Portable Handheld Battery powered, lightweight, easy to carry Home visits, field screening, mobile clinics
Desktop/Clinical Advanced features, computer connectivity, printable results Hospitals, audiology centers
High Frequency (1000 Hz) Specifically designed for infants under 6 months Newborn and infant testing
Note: Standard tympanometers use 226 Hz frequency for children over 6 months and adults. High frequency tympanometers (1000 Hz) are needed for infants under 6 months because their ear canals are different.

How to Use a Tympanometer: Step-by-Step Guide

Before Starting

  • Check that the device is clean and functioning properly
  • Ensure probe tips are clean and appropriate size
  • Have spare probe tips available
  • Explain the test to the patient (especially children)
  • Ensure the patient is calm and still

Testing Procedure

  1. Examine the ear: Use an otoscope to check for earwax blocking the ear canal. Remove excess wax if needed. Check for visible ear problems.
  2. Select probe tip: Choose the right size probe tip for the patient's ear canal. It should fit snugly but comfortably. For children, smaller tips are needed.
  3. Prepare the patient: Ask the patient to sit still and stay quiet. For young children, they may sit on a parent's lap. Explain they will feel a brief pressure sensation.
  4. Insert the probe: Gently pull the outer ear upward and backward (for adults) or downward and backward (for young children) to straighten the ear canal. Insert the probe tip to create a seal.
  5. Start the test: Press the button to begin. The device will automatically change air pressure in the ear canal. This takes 1 to 3 seconds. The patient should not talk, swallow, or move.
  6. Check the seal: If the test fails or shows an error, the seal may be poor. Adjust the probe position and retry. Ensure no air leaks around the probe.
  7. Record results: Note or print the tympanogram graph and measurements. Test both ears.
  8. Clean the probe: Remove the used probe tip. Clean or dispose of it according to infection control guidelines. Wipe the probe with appropriate disinfectant.

Understanding Results

Tympanometry results are shown as a graph with three main patterns:

Type Graph Pattern Meaning
Type A (Normal) Sharp peak at or near 0 pressure Normal middle ear function, normal eardrum movement
Type B (Flat) No clear peak, flat line Fluid in middle ear, blocked eustachian tube, or eardrum perforation
Type C Peak shifted to negative pressure Negative middle ear pressure, eustachian tube dysfunction
Important: Tympanometry results should always be interpreted along with other examinations and patient symptoms. A healthcare professional should make the final diagnosis.

Precautions and Safety

When NOT to Use

Do not perform tympanometry if:
  • Recent ear surgery (within 6 weeks)
  • Acute ear pain or suspected eardrum rupture
  • Drainage coming from the ear
  • Foreign object visible in the ear canal
  • Severe ear canal inflammation or infection

Safety Precautions

  • Always inspect the ear before testing
  • Use gentle pressure when inserting the probe
  • Never force the probe into the ear canal
  • Stop immediately if patient experiences pain
  • Ensure probe tips are clean and sterile
  • Use one-time disposable tips when possible
  • Follow infection control protocols strictly
  • Keep the device calibrated according to manufacturer guidelines

For Children

  • Explain the test in simple, child-friendly language
  • Allow parent to hold younger children during test
  • Use distraction techniques for anxious children
  • Choose appropriate probe tip size for small ear canals
  • Be extra gentle with insertion
  • Work quickly to minimize discomfort

Possible Discomfort

  • Brief feeling of fullness or pressure in the ear (normal)
  • Slight popping sensation (normal)
  • Temporary feeling of ear being blocked (resolves immediately)
  • No pain should occur during normal testing

Frequently Asked Questions

Is tympanometry painful?
No. Tympanometry is painless. You may feel brief pressure or fullness in the ear, similar to flying in an airplane, but it should not hurt.
How long does the test take?
Each ear takes about 1 to 3 seconds to test. The entire procedure including preparation usually takes 5 to 10 minutes for both ears.
Can babies have tympanometry?
Yes. Babies can be tested, but infants under 6 months need a special high frequency (1000 Hz) tympanometer for accurate results.
Does my child need to be still?
Yes. The child should remain still and quiet for the few seconds during testing. Movement, talking, or swallowing can affect results.
Will tympanometry detect hearing loss?
No. Tympanometry tests middle ear function, not hearing. A separate hearing test is needed to measure hearing ability.
What if there is earwax?
Excessive earwax can block the test. The ear canal should be checked and cleaned if needed before testing.
Can I have tympanometry with a cold?
Yes, but results may be abnormal due to temporary eustachian tube congestion. Your doctor may recommend retesting after the cold resolves.
How often should tympanometry be repeated?
This depends on the condition being monitored. Your healthcare provider will advise on follow-up testing schedule.
Are there any side effects?
No. Tympanometry has no side effects when performed correctly. The pressure sensation resolves immediately after testing.
Can I buy a tympanometer for home use?
Portable tympanometers are available, but proper training is needed to use them correctly and interpret results. Consult with a healthcare professional for proper diagnosis.

Device Maintenance and Care

Daily Care

  • Wipe the probe with alcohol wipes or recommended disinfectant after each use
  • Dispose of single-use probe tips properly
  • Check probe for damage or debris
  • Store device in clean, dry location
  • Keep away from extreme temperatures

Regular Maintenance

  • Calibrate device according to manufacturer schedule (usually every 6-12 months)
  • Replace probe tips regularly
  • Check batteries in portable units
  • Clean air tubes and filters as recommended
  • Update software if applicable
  • Keep device manual accessible

Storage Guidelines

  • Store in protective case when not in use
  • Keep in temperature controlled environment (usually 15-30 degrees Celsius)
  • Avoid direct sunlight
  • Protect from dust and moisture
  • Store probe tips in sterile packaging
  • Keep power cords organized and undamaged

Troubleshooting Common Issues

Problem Possible Solution
Test fails repeatedly Check probe seal, remove earwax, ensure patient is still, try different probe tip size
Inconsistent results Calibrate device, check for probe damage, ensure proper technique
No power Check batteries, verify power connection, check circuit breaker
Error messages Consult device manual, restart device, contact technical support
Professional Service: Have the device professionally serviced and calibrated annually or according to manufacturer recommendations. Do not attempt internal repairs yourself.

Additional Information

Training Requirements

Proper training is essential for accurate tympanometry. Healthcare professionals should receive instruction on:

  • Ear anatomy and physiology
  • Device operation and calibration
  • Proper testing technique
  • Result interpretation
  • Troubleshooting and maintenance
  • Infection control procedures

Quality Assurance

  • Perform daily equipment checks
  • Use test cavities to verify accuracy
  • Document calibration dates
  • Keep records of maintenance
  • Monitor result patterns for consistency
  • Report equipment malfunctions promptly

Infection Control

  • Use disposable probe tips whenever possible
  • Clean and disinfect reusable components between patients
  • Follow healthcare facility infection control protocols
  • Wash hands before and after testing
  • Use gloves when examining ears with drainage
  • Properly dispose of contaminated materials

Reference Resources

For more detailed information, consult:

  • American Academy of Audiology clinical practice guidelines
  • American Academy of Pediatrics ear examination resources
  • Device manufacturer operation manuals and training materials
  • Audiology textbooks on acoustic immittance testing
  • WHO guidelines on ear and hearing care
  • National hearing screening program guidelines in your country
Medical Disclaimer: This guide is for educational and informational purposes only and is not a substitute for professional medical advice, diagnosis, or treatment. Tympanometry should be performed by trained healthcare professionals. Always seek the advice of a qualified healthcare provider with any questions regarding medical conditions or test results. Never disregard professional medical advice or delay seeking it because of information on this page. The information provided here is accurate to the best of our knowledge but medical practices and guidelines may change. Always refer to current clinical guidelines and manufacturer instructions for your specific device.
Content checked and reviewed by a licensed pediatrician

Labels: